©Journal of Sports Science and Medicine (2013) 12, 414 - 421

Research article
Microcirculation Under an Elastic Bandage During Rest and Exercise - Preliminary Experience With the Laser-Doppler Spectrophotometry System O2C
Björn Sommer,1 , Gereon Berschin2, Hans-Martin Sommer2  
Author Information
1 University of Erlangen-Nuremberg, Erlangen, Germany
2 Philipps-University Marburg, Marburg, Germany

Björn Sommer
✉ Department of Neurosurgery, University of Erlangen-Nuremberg, Schwabachanlage 6, 91054 Erlangen, Germany
Email: bjoern.sommer@uk-erlangen.de
Publish Date
Received: 10-01-2013
Accepted: 05-04-2013
Published (online): 01-09-2013
 
ABSTRACT

There is an abundace of studies on the influence of rest and exercise as well as external compression on cutaneous, subcutaneous and muscle tissue blood flow using different measurement techniques. As a novel approach, we simultaneously examined the influence of a custom- made elastic thigh bandage on cutaneous and subcutaneous venous blood oxygenation (SO2), postcapillary venous filling pressures (rHb) and blood flow (flow) using the non-invasive laser- Doppler spectrophotometry system “Oxygen-to-see(O2C)”. Parameters were obtained in 20 healthy volunteers in 2 mm and 8 mm tissue depth during rest, 5 and 10 minutes of moderate bicycle exercise following a 10-minute recovery period. Without the bandage, results matched the known physiological changes indicating higher blood backflow from superficial and deep veins. Underneath the elastic bandage, we observed lower post-capillary filling pressures during exercise. However, after the bandage was removed in the post-exercise period, all obtained parameters of microcirculation remained increased, indicating a higher amount of local venous blood volume in this area. Our observations might be the result of external compression, thermoregulatory and exercise-dependent vascular mechanisms. With the O2C device, a promising new non- invasive technique of measuring local microcirculation in soft tissue exists. This study gives new insights in the field of non-invasive diagnostics with special regard to the influence of elastic bandages on local microcirculation.

Key words: External compression, blood flow, non-invasive diagnostics, lower extremity, exercise


           Key Points
  • It can be demonstrated that a novel non-invasive laser-Doppler spectrophotometry system allows the determination of capillary-venous microcirculation in an in-vivo study during exercise-rest cycles.
  • The results received with this technique indicate that a) without an elastic thigh bandage, turnover rates of capillary and post-capillary microperfusion in skin and subcutaneous fat tissue increase under physical exertion, b) skin blood flow decreases while subcutaneous blood flow remained constant in the subsequent recovery phase. While wearing the bandage, c) venous back flow during exercise is increased, whereas d) in the recovery phase, microcirculation remained increased in both tissue depths after removing the bandage.
  • In conclusion, the elastic bandage has a negative impact on local microcirculation and capillary-venous back flow, which is possibly due to a displacement of blood volume into the deep venous system and heat accumulation impairing the thermoregulatory response at the same time.

INTRODUCTION

Elastic bandages and compression stockings are widely used as a basic treatment of chronic venous insufficiency of the lower extremity and for the prevention of deep venous thrombosis as well as lymphedema (Aschwanden et al., 2008; Nehler et al., 1993). Both orthopaedic devices have a mutual mechanical and proprioceptive benefit on joint stabilization (Beynnon et al., 2002; Hassan et al., 2002). However, circular compression creates an elevated hydrostatic pressure in the tissue beneath the brace (Styf, 1999). As a consequence, there is an elevation of intramuscular pressure at rest and during exercise, which impairs muscle blood flow and transportation of metabolites leading to premature muscle fatigue (Man and Morrissey, 2005; Styf, 1990; Styf et al., 1992; 1994). Since many patients' microcirculation is impaired after knee or foot injury, external compression by elastic bandages and braces could have potential detrimental effects during the healing process.

Invasive and non-invasive methods of measuring cutaneous and subcutaneous tissue perfusion are well established. Venous occlusion plethysmography and radionuclide techniques are the most commonly used invasive methods that allow a quantitative description of local and regional blood flow (De Graaff et al., 2003; Sejrsen and Bülow, 2009; Martin et al., 2011). However, microtrauma can cause false positive and negative results and invasive measurements require an elaborate experimental setup. When using radionuclides, tracer biokinetics such as isotope affinity to fat tissue and diffusibility are known to bias blood flow values of the terminal capillary system (Swain and Grant, 1989). The high sensitivity to movement artifacts is one major problem of the assessment of regional blood flow using venous occlusion plethysmography during moderate upright exercise (Rowell, 1983). Laser-Doppler spectrophotometry is a promising new non-invasive tool which measures qualitative changes in capillary and venous microcirculation. The combination of backscattering-spectroscopy and laser-Doppler flowmetry allows real-time determination of venous oxygen saturation, haemoglobin saturation and blood flow. The monitoring device used in this study (“Oxygen-to- see(O2C)”, LEA Medizintechnik, Giessen, Germany), has been proven to be a reliable and valid tool in various conditions such as transplant and plastic surgery (Knobloch et al., 2003; Hölzle et al., 2010), Achilles tendon and ankle microcirculation (Knobloch et al., 2006a; Knobloch et al., 2006b), or wound tissue microperfusion in pressure and diabetic foot ulcers (Reenalda et al., 2009; Beckert et al., 2004).

In this study, the O2C device is used for the first time to evaluate local venous microcirculation of the thigh in skin and subcutaneous tissue depths. Apart from the physiological reactions to circulatory activity, the aim of this study was to examine if a custom- made thigh bandage had a positive or negative impact on soft tissue microperfusion during rest and exercise conditions in healthy volunteers.

METHODS

Subjects

Twenty healthy non- smoking male volunteers [mean ± SD: 27.1 ± 7.7 years] participated in this study. Regarding sports activity, 11 participants performed endurance training on a regular basis (7.6 ± 4.1 h per week) and 9 participants fitness training with power lifting (9.6 ± 1.7 h per week). All subjects were free of any injuries of the lower extremities and proved full functionality of the hip, knee and ankle joints under physical investigation. Anthropometric data for the subjects were: body mass 82. 1 ± 9.8 kg and height 183 ± 8 cm. The circumference of the thigh (measured 15 cm above the superior border of the patella) was 50.2 ± 2.8 cm.

This study complied to the laws of the Federal Republic of Germany and followed the Declaration of Helsinki. Subjects gave written informed consent before the tests.

Experimental setup

We used a modified open, non-blinded crossover-study design. Each participant completed two exercise trials, one with and one without wearing an elastic bandage. The trials were arranged in random order and at least 72 hours apart. External factors such as ambient room temperature (20.0 ± 1°C), air humidity (20%), air pressure (1013 hPa) and no direct illumination were kept constant in our laboratory.

At the beginning of each session the participants were seated in an upright position on the ergometer bicycle. The Oxygen-to-see probes were placed on the most distal part of the M. rectus femoris and gently strapped with tape (Leukoplast® hospital, BSN medical, Hamburg, Germany) as shown in Figure 1. Standardization of the probe position was achieved with additional reference to the bone structures (patella, medial and lateral femoral condyles). The tape application had no effect on the measured values.

The pre-exercise parameters (condition “pre-exercise”) venous oxygen saturation (SO2), postcapillary venous filling pressures (rHb) and microcirculatory blood flow (flow) were measured at two distinct tissue depths (2 mm and 8 mm) after resting 10 minutes on the ergometer bicycle. Pre- study examinations showed that after this period of time, all parameters remained in a steady condition. This measurement was made in both recording sets.

Thereafter, subjects started cycling in the first set, while in the second set the elastic bandage was adjusted. The custom-made bandage (width 13.7 cm, material: mixture of neoprene and polyurethane foam) was wrapped around the right thigh 15 cm above the superior border of the patella, a similar position to that of commonly used elastic knee bandages (e.g. Bauerfeind, 2011). The bandage could be adjusted to the individual size of the leg by tightening two straps with a defined tensile force of 25 N, which was measured by a tensiometer (Lehrmittelbau MAEY, Bonn, Germany). We chose the dimension of tensile force according to the results of Lundin and Styf, 1998, who discovered that the participants' preferred level of tightening the thigh straps of a knee brace showed similar effects on intramuscular pressure compared to a tensile force of 25 N.

During exercise, we performed real-time measurements after 5 and 10 minutes of cycling in both sets. In the last recording (“post-exercise”), exercise was stopped and subjects had to rest in an upright position on the ergometer for ten minutes. After this time span, parameters in participants without the bandage were measured in the first set, while in the second set, participants who wore the bandage had to remove it before parameters were obtained. Due to the known physiological changes in the vasomotor tone of skin and subcutaneous blood vessels during initiation of exercise, the first 5-10 minutes and the post-exercise period (Johnson, 1992; Roddie, 1983), we chose to focus our data collection to four distinct moments rather than using the complete data set.

Throughout the examination, the heart rate was transmitted to the ergometer (Lifecycle® 9500HR, LifeFitness, Brunswick Corporation, Schiller Park, Illinois, U.S.A.) by a chest strap (Polar T31®, Polar Electro Oy, Kempele, Finland). The ergometer used an electromagnetic brake system to adjust pedal resistance according to the individual training condition. Subjects had to perform the test with a steady heart rate of 130 bpm, which is considered as “moderate intensity” of endurance-type activity (Pollock et al., 1998).

Determination of blood circulation

Two flat probes were connected to the laser- Doppler spectrophotometry system “Oxygen-to-see (O2C)”, the first measuring the microperfusion to a depth of 2 mm, which represented cutaneous tissue (“superficial probe”), the other probe collected data to a depth of 8 mm, which represented subcutaneous tissue (“deep probe”). The white light source was calibrated with a definite raw haemoglobin spectrum before each measurement. All data were recorded in real-time at 50 Hz, which allowed pulsed synchronous measurements. The technology of this device combines two types of measuring components, backscattering-spectroscopy and laser-Doppler flowmetry. Since these optical methods have been described in detail elsewhere (Frank et al., 1989; Knobloch et al., 2003; 2006b), we give only a brief overview in the following passage.

Light that is sent through tissue will be reflected i.e. by blood cells at certain wavelengths, depending on their haemoglobin content and oxygen saturation (Figure 2). Through the specific absorption spectra it is possible to determine local oxygen supply parameters and the amount of local haemoglobin relative postcapillary filling pressures by a spectrophotometer (Knobloch et al., 2006b). At the same time, the integrated laser-Doppler flowmeter uses the Doppler principle to measure the relative velocity of moving particles. Here, laser light of 830 nm wavelength is sent through a fiber-optic cable. If photons are backscattered by moving blood cells, they undergo a frequency shift, which is collected by the same probe (Eyre et al., 1988; Bonner and Nossal, 1990). Using the Doppler frequency shift formula, the velocity of blood cells and, as a combination of velocity and quantity, relative blood flow can be determined (Michelson et al., 1996; Seifalian et al., 1994).

The O2C system uses arbitrary units (“AU”) to describe the quantity of haemoglobin amount. The manufacturer introduced this unit because measured values are derived from electrical units. To calculate the blood flow in millilitres per minute, the electrical signals have to be compared to a corresponding method, i.e. coloured microspheres (Walter et al., 2002) and have to be calibrated for each measured organ. Measurements with the O2C system reflect mainly the capillary-venous compartment of the microcirculation, because about 80% of the haemoglobin is located in the capillary and post-capillary system of the microvascular bed (Gaethgens, 2000; Abboud et al., 1976). Since light is absorbed completely if the vessel diameter is greater than 100µm, the O2C system measures only the blood parameters in nutritive, microvascular vessels (Gandjbakhche et al., 1999).

Since the capillary-venous vessel bed is serving as a volumetric store, it guarantees that the same amount of blood volume is transported through the venous system at low pressure as it is in the corresponding high pressure arterial system in order to maintain blood flow (Sinaasappel et al., 1999). Under physiological conditions, a rise of venous oxygen saturation and microcirculatory blood flow is expected in both tissue depths during exercise, while postcapillary venous filling pressures decrease due to higher venous outflow. An inhibited venous capillary outflow would produce a decrease in SO2 and flow with an increase of rHb (Figure 3), because of the accumulation of blood volume through insufficient blood transportation to the deep veins (Hanna et al., 1997; Hölzle et al., 2010; LEA, 2007). The reliability and intrasubject variability of the laser-Doppler system has been examined in several studies and proved sufficient under standardized testing conditions (Brell et al., 2005; Ghazanfari et al., 2002; Walter et al., 2002; Wunder et al., 2005).

Statistical analyses

Results are given as mean values and one standard deviation (SD). With a sampling rate of 0.5 per second of the O2C device, we calculated the average from 15 sample values over 30 seconds for each of the four recordings. Changes were assessed by a two-way analysis repeated measures analysis of variance (ANOVA) with time (“pre-exercise, 5 minutes, 10 minutes, post-exercise”) and condition (with/without bandage) as factors. To meet the requirements for a normal distribution, we examined each variable independently by a KOLMOGOROV-SMIRNOV-Test. For homogenity of variances, a Levene-test was used. Bonferroni's post-hoc tests for multiple comparisons were used when differences reached statistical significance. The average intrasubject variability of the blood flow parameters is 5 %, whereas the reproducibility of these values did not show significant changes within or between different days (Ghazanfari et al., 2002). Significance was accepted when P was less than 0.05. The SPSS statistical software package 15.0 for Windows (SPSS Inc., Chicago, Illinois) was used for statistical analysis.

RESULTS

Results are shown in Tables 1 and 2">2. In summary, we made the following observations:

We compared the results of both sessions (“inter-subject differences”) and found a statistically significant difference between the venous blood amount (rHb) in both skin and subcutaneous tissue. While subjects were wearing the elastic bandage, rHb was lower in skin (Table 2) and subcutaneous tissue (Table 2) during pre-exercise measurements. After 5 minutes of cycling, skin rHb was 59.3 ± 20.4 AU compared to 66.5 ± 8.8 AU without the bandage (p = 0.009), whereas there were no significant differences in 8 mm detection depth (p = 0.093). Besides that, cutaneous flow was significantly (p = 0.001) higher underneath the bandage (Table 2) in the post-exercise period, whereas the corresponding values showed no difference in subcutaneous tissue (Table 2). In all other measurements, no difference could be detected.

DISCUSSION

Many studies exist that describe the influence of external compression on cutaneous, subcutaneous and muscle tissue blood flow using non-invasive measurement techniques. However, most of them have chosen a reduced experimental setup where “physical activity” is defined as active or passive movement (flexion/extension) of unilateral ankle joints. The obvious methodological consideration is avoiding side effects and create a simplified, standardized experimental setting (Pearson et al., 2011; Stein et al., 2010; Zelis et al., 1969; Zhang et al., 2004), although this does not fully reflect the actual complexity of the influence of circulatory and thermoregulatory mechanisms on local microcirculation, as seen in a more “physiological” type of physical activity such as running or cycling.

Moreover, non-invasive studies that deal with the influence of external compression on skin blood perfusion in rest and exercise conditions are rare (Mayrovitz and Sims, 2003; Styf, 1999; Styf et al., 1992). By trying to avoid side-effects of placing the probe between the bandage-skin interface, some studies performed laser-Doppler investigations beneath the bandage (Mayrovitz et al., 1998) or used single-point laser-Doppler flowmetry (Melhuish et al., 2004) to measure microcirculation under a bandage without removing it. However, in both studies, the results were based on the recording of microcirculatory blood flow only with no additional information about the oxygenation or blood volume level. Besides, the circulatory provocation methods were simply defined as an increase in external pressure applied to the calf by an air cast (Mayrovitz et al., 1998).

To the best of our knowledge, no previous study has measured local microcirculation in two different tissue depths using a laser-Doppler spectrophotometer and determined the effects of external compression caused by an elastic bandage on blood perfusion under endurance-type activity.

Our first results when no elastic bandage was worn matched the expected physiological changes. We observed a rise of venous blood oxygenation (SO2) and blood flow (flow) in cutaneous and subcutaneous tissue, whereas the blood amount indicated by postcapillary venous filling pressures (rHb) was reduced during exercise. Blumberg and colleagues (2005) measured with the O2C system at 2, 8 and 16 mm detection depth while performing a contralateral handgrip exercise. After 10-12 seconds, blood flow increased in all tissues with a clear enhancement on the measurements obtained in skin at 2 mm tissue depth. In our opinion, this is an expression of higher muscle blood flow and increased turnover rates of capillary and post-capillary microperfusion, which leads to a higher venous return to the right ventricle from superficial and deep veins.

Our post-exercise results revealed a reduction in skin blood flow by 31% (from 313.2 to 216.1 AU, p < 0.001), while SO2 had a tendency to decrease and rHb to increase. In 8 mm tissue depth, all parameters remained constant. This could be a result of the thermoregulatory reflexes mediating a stepwise elimination of thermal energy and metabolites generated by muscular activity.

While subjects were wearing the elastic bandage in the second set, we observed a tendency of higher venous backflow during exercise expressed as lower post-capillary filling pressures. These differences were statistically significant in skin tissue only and lasted throughout the bicycle exercise period. Changes in the obtained microcirculatory parameters indicate that the amount of venous blood volume in 2 mm and 8 mm depth is less while wearing an elastic bandage during rest and exercise conditions. However, after the bandage was removed in the post-exercise period, all parameters remained increased instead of showing the expected changes (i.e. reduction in SO2 and flow, elevation in rHb) as one would see in a “physiological” condition without wearing a bandage. Therefore, we conclude that the amount of venous blood volume must be higher under elastic bandages. This is contrary to the above mentioned reduction of venous blood volume in both tissue depths, but can possibly be explained by the following mechanisms:

Thus, we conclude that wearing an elastic thigh bandage has an impact on venous blood flow in cutaneous and subcutaneous tissue depths. Our results suggest that there is a higher amount of venous blood volume shift caused by application of an elastic bandage compared to physiological conditions without a bandage.

Study limitations

Limitations of our study are the non-invasive measurement technique and the lack of intramuscular pressure recordings. Our data was collected in two distinct tissue depths: 2 mm (cutaneous layer) and 8 mm (subcutaneous fat layer). Depending on age, weight and physical condition, average subcutaneous fat thickness of the quadriceps muscle in males is between 4-15 mm (Maurits et al., 2003). With a skinfold thickness of 7.1 ± 2.9 mm at the measurement site, part of the signal from the subcutaneous probe not only resembled subcutaneous, but also muscular blood flow from M. rectus femoris. Moreover, subcutaneous tissue oxygenation corresponds directly to the viability of the adjacent muscle tissue and gives an indirect measure of muscle blood flow. This could be the reason that we observed a higher variance expressed as a large standard deviation in flow only in subcutaneous tissue depth throughout this study.

CONCLUSION

In our study, we were the first to record microcirculatory parameters in skin and subcutaneous tissue depths with the non-invasive combined backscattering-spectroscopy and laser-Doppler flowmetry device “Oxygen-to-see” in healthy subjects under rest and exercise conditions. In the first test regime, we observed a rise in venous oxygenation and blood flow along with a small decrease in venous blood amount as an expression of higher muscular blood flow and venous backflow to the right ventricle from superficial and deep veins. In the second set, we noticed lower post-capillary filling pressures while subjects were wearing a unilateral thigh bandage. This trend might be the result of an increase in venous back flow. However, in the post-exercise period, all parameters remained increased compared to the physiological condition without a bandage, indicating that external compression and thermal isolation caused by the bandage could generate venous blood pooling in both tissue depths. Using a realistic and not simplified laboratory setup, the results of this study underline the possibilities of non-invasive monitoring with a novel technique and the impact of an elastic thigh bandage on skin and subcutaneous microcirculation. In the future, further controlled studies are needed to evaluate and compare these findings with “gold standard” invasive diagnostic methods.

ACKNOWLEDGEMENTS

We gratefully acknowledge the support of Mrs. Verena Schneider, MEd, who assisted in the acquisition of data.

AUTHOR BIOGRAPHY

Journal of Sports Science and Medicine Björn Sommer
Employment: Resident physician at the Department of Neurosurgery, University of Erlangen-Nuremberg, Erlangen, Germany.
Degree: MD
Research interests: Non-invasive diagnostics, Orthotic management of the lower limb, Haemodynamics
E-mail: bjoern.sommer@uk-erlangen.de
 

Journal of Sports Science and Medicine Gereon Berschin
Employment: PD at the Department of Sports Science and Motology, Philipps-University Marburg, Marburg, Germany
Degree: PhD
Research interests: Exercise and Training, Biomechanics, Posture, Rehabilitation
E-mail: berschin@staff.uni-marburg.de
 

Journal of Sports Science and Medicine Hans-Martin Sommer
Employment: Professor emeritus of the Department of Sports Science and Motology, Philipps-University Marburg, Marburg, Germany
Degree: MD
Research interests: Motor control, Gait and Posture, Pathomechanisms of micro- and macrotraumata in sports
E-mail: hmsommer@t-online.de
 
 
REFERENCES
Journal of Sports Science and Medicine Abboud F.M., Mark A.L., Heistad D.D., Schmid P.G., Barnes R.W., Levine H. (1976) Cardiovascular pathophysiology. The venous system. New York, U.S.A.. Grune & Stratton.
Journal of Sports Science and Medicine Aschwanden M., Jeanneret C., Koller M.T., Thalhammer C., Bucher H.C., Jaeger K.A. (2008) Effect of prolonged treatment with compression stockings to prevent post-thrombotic sequelae: a randomized controlled trial. Journal of Vascular Surgery 47, 1015-1021.
Journal of Sports Science and Medicine Bauerfeind AG (2011) Product catalog 02/2011: Supports and orthoses. Zeulenroda-Triebes, Germany.
Journal of Sports Science and Medicine Beckert S., Witte M.B., Königsrainer A., Coerper S. (2004) The impact of the Micro-Lightguide O2C for the quantification of tissue ischemia in diabetic foot ulcers. Diabetes Care 27, 2863-2867.
Journal of Sports Science and Medicine Beynnon B.D., Good L., Risberg M.A. (2002) The effect of bracing on proprioception of knees with anterior cruciate ligament injury. Journal of Orthopaedic and Sports Physical Therapy 32, 11-15.
Journal of Sports Science and Medicine Blumberg H., Frisch S., Husmann J., Magerl W., Schleinzer W., Tiede W., Villiger D. (2005) 11th World Congress on Pain. A new approach to simultaneously monitor (sympathetic) vasomotor reactions of skin and underlying deep tissues in man. Sydney, Australia. Poster.
Journal of Sports Science and Medicine Bonner R.F., Nossal R., Shepherd A.P., Öberg P.Å. (1990) Developments in Cardiovascular Medicine: Laser-Doppler Blood Flowmetry. Principles of laser-Doppler flowmetry. Boston, U.S.A.. Kluwer Academic Publishers.
Journal of Sports Science and Medicine Brell B., Temmesfeld-Wollbrück B., Altzschner I., Frisch E., Schmeck B., Hocke A.C., Suttorp N., Hippenstiel S. (2005) Adrenomedullin reduces Staphylococcus aureus alphatoxin-induced rat ileum microcirculatory damage. Critical Care Medicine 33, 819-826.
Journal of Sports Science and Medicine Buckey J.C., Peshock R.M., Blomqvist C.G. (1988) Deep venous contribution to hydrostatic blood volume change in the human leg. American Journal of Cardiology 62, 449-453.
Journal of Sports Science and Medicine Dai X.Q., Lu Y.H., Lin H., Bai L. (2012) Mechanisms of control of human skin blood flow under external pressure. Biological Rhythm Research 43, 267-278.
Journal of Sports Science and Medicine De Graaff J.C., Ubbink D.T., Lagarde S.M., Jacobs M.J. (2003) Postural changes in capillary pressure in the hallux of healthy volunteers. Journal of Applied Physiology 95, 2223-2228.
Journal of Sports Science and Medicine Eyre J.A., Essex T.J., Flecknell P.A., Bartholomew P.H., Sinclair J.I. (1988) A comparison of measurements of cerebral blood flow in the rabbit using laser Doppler spectroscopy and radionuclide labelled microspheres. Clinical Physics and Physiological Measurement 9, 65-74.
Journal of Sports Science and Medicine Frank K.H., Kessler M., Appelbaum K., Dümmler W. (1989) The Erlangen micro-lightguide spectrophotometer EMPHO I. Physics in Medicine and Biology 34, 1883-1900.
Journal of Sports Science and Medicine Fromy B., Legrand M.S., Abraham P., Leftheriotis G., Cales P., Saumet J.L. (1997) Effects of positive pressure on both femoral venous and arterial blood velocities and the cutaneous microcirculation of the forefoot. Cardiovascular Research 36, 372-376.
Journal of Sports Science and Medicine Gaethgens P., Klinke R., Silbernagl S. (2000) Lehrbuch der Physiologie. Das Kreislaufsystem. Stuttgart, Germany and New York, U.S.A.. Georg Thieme Verlag.
Journal of Sports Science and Medicine Gandjbakhche A.H., Bonner R.F., Arai A.E., Balaban R.S. (1999) Visible-light photon migration through myocardium in vivo. American Journal of Physiology 277, H698-704.
Journal of Sports Science and Medicine Ghazanfari M., Vogt L., Banzer W., Rhodius U. (2002) Reproducibility of non-invasive blood flow measurements using laser Doppler spectroscopy. Physikalische Medizin, Rehabilitationsmedizin, Kurortmedizin 12, 330-336.
Journal of Sports Science and Medicine Hanna G.B., Newton D.J., Harrison D.K., McCollum P.T. (1997) Use of lightguide spectrophotometry to investigate the effect of postural changes on skin oxygenation in deep venous insufficiency. British Journal of Surgery 84, 520-523.
Journal of Sports Science and Medicine Hassan B.S., Mockett S., Doherty M. (2002) Influence of elastic bandage on knee pain, proprioception, and postural sway in subjects with knee osteoarthritis. Annals of the Rheumatic Diseases 61, 24-28.
Journal of Sports Science and Medicine Hölzle F., Rau A., Loeffelbein D.J., Mücke T., Kesting M.R., Wolff K.D. (2010) Results of monitoring fasciocutaneous, myocutaneous, osteocutaneous and perforator flaps: 4-year experience with 166 cases. International Journal of Oral and Maxillofacial Surgery 39, 21-28.
Journal of Sports Science and Medicine Heinonen I., Brothers R.M., Kemppainen J., Knuuti J., Kalliokoski K.K., Crandall C.G. (2011) Local heating, but not indirect whole body heating, increases human skeletal muscle blood flow. Journal of Applied Physiology 111, 818-824.
Journal of Sports Science and Medicine Johnson J.M. (1992) Exercise and the cutaneous circulation. Exercise and Sport Sciences Reviews 20, 59-97.
Journal of Sports Science and Medicine Knobloch K., Kraemer R., Lichtenberg A., Jagodzinski M., Gosling T., Richter M., Krettek C. (2006b) Microcirculation of the ankle after Cryo/Cuff application in healthy volunteers. International Journal of Sports Medicine 27, 250-255.
Journal of Sports Science and Medicine Knobloch K., Kraemer R., Lichtenberg A., Jagodzinski M., Gossling T., Richter M., Zeichen J., Hufner T., Krettek C. (2006a) Achilles tendon and paratendon microcirculation in midportion and insertional tendinopathy in athletes. American Journal of Sports Medicine 34, 92-97.
Journal of Sports Science and Medicine Knobloch K., Lichtenberg A., Pichlmaier M., Mertsching H., Krug A., Klima U., Haverich A. (2003) Microcirculation of the sternum following harvesting of the left internal mammary artery. The Thoracic and Cardiovascular Surgeon 51, 255-259.
Journal of Sports Science and Medicine Lundin O., Styf J.R. (1998) Intramuscular pressure in the leg and thigh related to tensile strap force during knee brace wear. An experimental study in man. American Journal of Sports Medicine 26, 567-570.
Journal of Sports Science and Medicine Man I.O., Morrissey M.C. (2005) Relationship between ankle-foot swelling and self-assessed function after ankle sprain. Medicine and Science in Sports and Exercise 37, 360-363.
Journal of Sports Science and Medicine Martin E., Brassard P., Gagnon-Auger M., Yale P., Carpentier A.C., Ardilouze J.L. (2011) Subcutaneous adipose tissue metabolism and pharmacology: a new investigative technique. Canadian Journal of Physiology and Pharmacology 89, 383-391.
Journal of Sports Science and Medicine Maurits N.M., Bollen A.E., Windhausen A., De Jager A.E., Van Der Hoeven J.H. (2003) Muscle ultrasound analysis: normal values and differentiation between myopathies and neuropathies. Ultrasound in Medicine & Biology 29, 215-225.
Journal of Sports Science and Medicine Mayrovitz H.N., Delgado M., Smith J. (1998) Compression bandaging effects on lower extremity peripheral and sub-bandage skin blood perfusion. Ostomy/Wound Management 44, 56-60.
Journal of Sports Science and Medicine Mayrovitz H.N., Sims N. (2003) Effects of ankle-to-knee external pressures on skin blood perfusion under and distal to compression. Advances in Skin & Wound Care 16, 198-202.
Journal of Sports Science and Medicine Melhuish J.M., Krishnamoorthy L., Bethaves T., Clark M., Williams R.J., Harding K.G. (2004) Measurement of the skin microcirculation through intact bandages using laser Doppler flowmetry. Medical & Biological Engineering & Computing 42, 259-263.
Journal of Sports Science and Medicine Michelson G., Schmauss B., Langhans M.J., Harazny J., Groh M.J. (1996) Principle, validity, and reliability of scanning laser Doppler flowmetry. Journal of Glaucoma 5, 99-105.
Journal of Sports Science and Medicine Nehler M.R., Moneta G.L., Woodard D.M., Defrang R.D., Harker C.T., Taylor L.M. Jr., Porter J.M. (1993) Perimalleolar subcutaneous tissue pressure effects of elastic compression stockings. Journal of Vascular Surgery 18, 783-788.
Journal of Sports Science and Medicine Oldfield M.A., Brown M.D. (2006) Evaluation of the time course of vascular responses to venous congestion in the human lower limb. Journal of Vascular Research 43, 166-174.
Journal of Sports Science and Medicine Pearson J., Low D.A., Stöhr E., Kalsi K., Ali L., Barker H., González-Alonso J. (2011) Hemodynamic responses to heat stress in the resting and exercising human leg: insight into the effect of temperature on skeletal muscle blood flow. American Journal of Physiology. Regulatory, Integrative and Comparative Physiology 300, R663-673.
Journal of Sports Science and Medicine Pollock M., Gaesser G.A., Butcher J.D., Després J.P., Dishman R.K., Franklin B.A., Garber C.E. (1998) American College of Sports Medicine position stand: The recommended quantity and quality of exercise for developing and maintaining cardiorespiratory and muscular fitness, and flexibility in healthy adults. Medicine and Science in Sports and Exercise 30, 975-991.
Journal of Sports Science and Medicine Ramelet AA. (2002) Compression therapy. Dermatologic Surgery 28, 6-10.
Journal of Sports Science and Medicine Reenalda J., Van Geffen P., Nederhand M., Jannink M., Ijzerman M., Rietmann H. (2009) Analysis of healthy sitting behaviour: interface pressure distribution and subcutaneous tissue oxygenation. Journal of Rehabilitation Research and Development 46, 577-586.
Journal of Sports Science and Medicine Roddie I.C., Sheperd J.T., Abboud F.M. (1983) Handbook of Physiology. Circulation to skin and adipose tissue. Bethesda, U.S.A.. American Physiological Society.
Journal of Sports Science and Medicine Rowell L.B., Sheperd J.T., Abboud F.M. (1983) Handbook of Physiology. Cardiovascular adjustments to thermal stress. Bethesda, U.S.A.. American Physiological Society.
Journal of Sports Science and Medicine Seifalian A.M., Stansby G., Jackson A., Howell K., Hamilton G. (1994) Comparison of laser Doppler perfusion imaging, laser Doppler flowmetry and thermographic imaging for assessment of blood flow in human skin. European Journal of Vascular Surgery 8, 65-69.
Journal of Sports Science and Medicine Sejrsen P., Bülow J. (2009) Blood flow rate measurements with indicator techniques revisited. Clinical Physiology and Functional Imaging 29, 385-391.
Journal of Sports Science and Medicine Sinaasappel M., van Iterson M., Ince C. (1999) Microvascular oxygen pressure in the pig intestine during haemorrhagic shock and resuscitation. Journal of Physiology 514, 245-253.
Journal of Sports Science and Medicine Stein P.D., Matta F., Yaekoub A.Y., Ahsan S.T., Badshah A., Younas F., Denier J.E. (2010) Effect of compression stockings on venous blood velocity and blood flow. Thrombosis and Haemostasis 103, 138-144.
Journal of Sports Science and Medicine Styf J.R. (1990) The influence of external compression on muscle blood flow during exercise. American Journal of Sports Medicine 18, 92-95.
Journal of Sports Science and Medicine Styf J.R. (1999) The effects of functional knee bracing on muscle function and performance. Sports Medicine 28, 77-81.
Journal of Sports Science and Medicine Styf J.R., Lundin O., Gershuni D.H. (1994) Effects of a functional knee brace on leg muscle function. American Journal of Sports Medicine 22, 830-834.
Journal of Sports Science and Medicine Styf J.R., Nakhostine M., Gershuni D.H. (1992) Functional knee braces increase intramuscular pressures in the anterior compartment of the leg. American Journal of Sports Medicine 20, 46-49.
Journal of Sports Science and Medicine Swain I.D., Grant L.J. (1989) Methods of measuring skin blood flow. Physics in Medicine and Biology 34, 151-175.
Journal of Sports Science and Medicine Walter B., Bauer R., Krug A., Derfuss T., Traichel F., Sommer N. (2002) Simultaneous measurement of local cortical blood flow and tissue oxygen saturation by Near infra-red Laser Doppler flowmetry and remission spectroscopy in the pig brain. Acta Neurochirurgica. Supplement 81, 197-199.
Journal of Sports Science and Medicine Wunder C., Brock R.W., Krug A., Roewer N., Eichelbrönner O. (2005) A remission spectroscopy system for in vivo monitoring of hemoglobin oxygen saturation in murine hepatic sinusoids, in early systemic inflammation. Comparative Hepatology 4, 1-.
Journal of Sports Science and Medicine Zelis R., Mason D.T., Braunwald E. (1969) Partition of blood flow to the cutaneous and muscular beds of the forearm at rest and during leg exercise in normal subjects and in patients with heart failure. Circulation Research 24, 799-806.
Journal of Sports Science and Medicine Zhang Q., Andersson G., Lindberg L.G., Styf J. (2004) Muscle blood flow in response to concentric muscular activity vs. passive venous compression. Acta Physiologica Scandinavica 180, 57-62.
 
 
 
Home Issues About Authors
Contact Current Editorial board Authors instructions
Email alerts In Press Mission For Reviewers
Archive Scope
Supplements Statistics
Most Read Articles
  Most Cited Articles
 
  
 
JSSM | Copyright 2001-2024 | All rights reserved. | LEGAL NOTICES | Publisher

It is forbidden the total or partial reproduction of this web site and the published materials, the treatment of its database, any kind of transition and for any means, either electronic, mechanic or other methods, without the previous written permission of the JSSM.

This work is licensed under a Creative Commons License Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.